Product Features
Automatic External Defibrillator features include:
Three-step defibrillation process
Two-button operation
Extensive voice and visual prompts for the operator
Biphasic energy output
Lock-out protection to prevent inadvertent defibrillation
Device has an internal memory(60 hours)
ECG signals and events can be transferred to the computer.
Defi5 AED is provided with carrying case.
Device has a carrying handle for easy carrying.
It has long-life disposable LiMnO2 type battery. Battery life of at least 5 years. and Defi5 AED can work also with rechargeable battery.
AED Storage Software (Option)
Its used to assist upper supervisory system of Automatic External Defibrillator DEFI and support transmission of storage data from lower control system of DEFI. It is used for event documentation display of real-time ECG voice prompts waveform display of voice prompts and print of selected waveforms.
Product Specifications
Physical
Category |
Specification |
Dimensions |
303 x 216 x 89 mm |
Weight |
2.0 kg |
Operating Temperature |
0℃ to 40℃ |
Operating Humidity |
Relative humidity between 30% and 95% (non-condensing) |
Storage Temperature
(without battery) |
-20℃ to 55℃ |
Storage Humidity
(without battery) |
Up to 93% (non-condensing) |
|
|
Electrical Isolation
Category |
Specification |
Power |
Unit operates on internal battery only |
External Electrical Connections |
No external devices are attached to the unit |
Risk Current Category |
Internally powered equipment with defibrillator-proof BF type patient applied part (as per definition of IEC 60601-1 standard) |
|
|
Defibrillator
Specification |
Biphasic Truncated exponential |
150 150 200J |
10 sec. to 150J
15 sec. to 200 J |
9 sec. |
Less than 30 seconds |
Less than 35 seconds |
Less than 25 seconds |
22 audible prompts |
LED prompts |
Two buttons - On/Off Shock |
±15% into any impedance from 25 to 175Ω |
20Ω to 200Ω |
Waveform details
The table below provides details of the biphasic truncated exponential waveform delivered by the DEFI5 (set to 200J) when connected to resistive loads of 25 through 175 Ohms. The waveforms are characterized by typical values for peak current (Ip) duration of the first output phase and duration of the second output phase. Values are within 10%.
Output Energy
(J) |
Patient Impedance
(Ω) |
Ip1
(Amps) |
Ip2
(Amps) |
Phase1
(ms) |
Phase2
(ms) |
Interval
(ms) |
200 |
25 |
53.6 |
36.0 |
5.5 |
3.3 |
0.7 |
50 |
27.8 |
17.5 |
10.9 |
6.6 |
0.7 |
75 |
18.2 |
12.0 |
16.3 |
8.9 |
0.7 |
100 |
13.4 |
9.0 |
19.7 |
9.6 |
0.7 |
125 |
10.5 |
7.2 |
20.5 |
11.2 |
0.7 |
150 |
8.6 |
6.0 |
21.6 |
10.4 |
0.7 |
175 |
8.2 |
6.1 |
21.8 |
11.2 |
0.7 |
Related Accessories
Standard:
AED Pad .................... 1Pcs
Battery ...........................1Pcs
AED Case ...................... 1Pcs
Cleaning Cloth ................1Pcs
Options:
AED Pad for Child
AED Cabinet
IrDA USB Adapter for ECG Data transfer CD Software for ECG Data transfer