Product Features
Audible and visual alarm for occlusion empty near empty low battery end of infusion syringe loose wrong setting etc
LCD Display high capacity words friendly user interface dynamically display working status
Compatible with 10ml 20ml 30ml 50/60ml syringe of any brands
Preset solution volume to greatly reduce the workload of nurses
Three work modes: Rate mode Time Volume mode Dosage Weight mode
Three levels of occlusion: high medium and low
Purge and Bolus function
KVO (keep-vein-open) automatically opens as injection is completed KVO rate is 1ml/h also can be modified from 0.1ml/h to 10ml/h
Freely Stackable: freely stack one syringe pump onto another to provide multiple solutions which have a wide range of clinical applications.
Connecting Port for Nurse-Paging
Power Source: AC100-240V 50/60Hz; Internal Battery DC11.1V car charge
One-key operation makes setup easy and simple
Syringe Plunger Grabble Detector operable with one hand in a No-Germ Environment
Automatically record the settings for last infusion(MD910s have 2000 records)
Drug library is available according to customers’ request(MD910s)
Product Specifications
Rate:
50/60ml Syringe: 0.1ml/h~999.9ml/h (0.1ml/h step)
1000ml/h~1500ml/h (1ml/h step)
30ml Syringe: 0.1ml/h~900.0ml/h (0.1ml/h step)
20ml Syringe: 0.1ml/h~600.0ml/h (0.1ml/h step)
10ml Syringe: 0.1ml/h~300.0ml/h (0.1ml/h step)
Flow Rate Accuracy: Within ±3% (after correct calibration)
Mechanical precision: Within ±2%
Bolus Rate:
50/60ml Syringe: 1200ml/h
30 ml Syringe: 720ml/h
20 ml Syringe: 480ml/h
10 ml Syringe: 240ml/h
Purge Rate: 50/60ml Syringe: 1500ml/h
30ml Syringe: 900ml/h
20ml Syringe: 600ml/h
10ml Syringe: 300ml/h
Volume Limit: 0.1ml~999.9ml
Total Injection Volume: 0.1ml~9999.9ml (0.1ml step)
Occlusion: High: 800mmHg ±200mmHg (106.7kPa±26.7kPa)
Medium: 500mmHg ±100mmHg (66.7kPa±13.3kPa)
Low: 300mmHg ±100mmHg (40.7kPa±13.3kPa )
Alarms: Injection soon finish end of injection occlusion improperly installation of syringe wrong setting low battery syringe loose etc.
Power Source: AC 100V~240V 50/60Hz; Internal rechargeable Li battery capacity≥1600mAh 4 hours internal battery backup
KVO rate: 1ml/h can be modified from 0.1ml/h to 10ml/h
Fuse: F1AL/250/60V 2pcs inside
Power Consumption: 18VA
IP Classification: IPX1
Equipment Classification: Class II internal power supply Type BF
Operating Condition: Ambient temperature: +5℃~+40℃;
Relative humidity: 20~90%
Transport & Storage Condition: Ambient temperature: -30℃~+55℃
Relative humidity: ≤95%
Dimension: 280mm (L) × 210mm (W) × 130mm (H)
Weight: 2.2kg (net weight)
Note:
Disposable syringe and extension tube sterilized by ethylene oxide that used with this product shall comply with the standards of GB15810-2001 sterile hypodermic syringes for single use. Syringe or extension tubes which do not comply with the standards may lead to incorrect injection rate drug residues and other possible hazard.