MD908C
Patient monitor 12inch TFT. Automatic temperature drift treatment of NIBP measuring patented technology;Friendly and configurable user interface;Powerful data review:Maximum 720-hour graphic and tabular trends,
Up to10000 pieces of NIBP record storage,
Wired or wireless networkable with Central Monitoring System;On-line upgrading keeps you updated;
Price: 210 USD
Buy 25: - 20% OFF
Up to10000 pieces of NIBP record storage,
Wired or wireless networkable with Central Monitoring System;On-line upgrading keeps you updated;
Price: 210 USD
Buy 25: - 20% OFF
Product Features
High resolution 12” color TFT display
12 Channel waves.9 Parameters : ECG Resp NIBP SpO2 TEMP
Light weight compact and portable
Built-in rechargeable lithium battery
Unique human voice alarm
Rich screen layouts: Standard Big font Trend table NIBP trend ECG full lead OxyCRG IoC Layout
Powerful storage: Maximum 720 Hour graphic and tabular trends of all
parameters up to 10000 NIBP review 200alarm review 2Hours of all waveforms
System software can be upgraded through network
Data and waveforms color be adjustable
Arrhythmia analysis and S-T segment analysis
Suitable for adult paediatric and neonatal patient
Options
Invasive Blood Pressure (IBP)
Carbon dioxide (CO2) Module
Thermal Printer
Touch Screen
Product Specifications
ECG
Lead Mode 5 Leads ( R L F N C or RA LA LL RLV)
Lead selection I II III avR avL avF V
Waveform 2 ch
Lead mode 3 Leads ( R L F or RA LA LL)
Lead selection I II III
Waveform 1 ch
Gain x2.5mm/mV x5.0mm/mV x10mm/mV x20mm/mV auto
HR and Alarm
Range
Adult 15 ~ 300 bpm
Neo/Ped 15 ~ 350 bpm
Accuracy ± 1% or ± 1bpmwhich great
Resolution 1 bpm
Sensitivity > 200 (uV )
Differential Input Impedance > 5 MΩ
CMRR
Monitor > 105 dB
Operation > 105 dB
Diagnosis > 85 dB
Electrode offset potential ±300mV
Leakage Current < 10 uA
baseline Recovery < 3 S After Defi.
ECG Signal Range ±8 m V ( Vp-p )
Bandwidth
Surgery 1 ~ 15 Hz
Monitor 0.5 ~ 35 Hz
Diagnostic 0.05 ~ 100 Hz
Calibration Signal 1 (mV ) Accuracy :±5%
ST Segment Monitoring Range
Measure and Alarm -2.0 ~ +2.0 mV
ARR Detecting
Type ASYSTOLE VFIB/VTAC COUPLET BIGEMINY TRIGEMINY R ON T VT>2 PVC TACHY BRADYMISSED BEATS PNP PNC
Alarm Available
Review Available
RESPARATION
Method Impedance between R-F(RA-LL)
Differential Input Impedance >2.5 MΩ
Measuring Impedance Range: 0.3~5.0Ω
base line Impedance Range: 0 – 2.5 KΩ
Bandwidth 0.3 ~ 2.5 Hz
Resp. Rate
Measuring and Alarm Range
Adult 0 ~ 120 rpm
Neo/Ped 0 ~ 150 rpm
Resolution 1 rpm
Accuracy ±2 rpm
Apean Alarm 10 ~ 40 S
NIBP
Method Oscillometric
Mode Manual Auto STAT
Measuring Interval in AUTO Mode 1 2 3 4 5 10 15 30 60 90 120 180 240480 (Min)
Measuring Period in STAT Mode 5 Min
Pulse Rate Range 40 ~ 240 bpm
Alarm Type SYS DIA MEAN
Measuring and alarm range
Adult Mode
SYS 40 ~ 270 mmHg
DIA 10 ~ 215 mmHg
MEAN 20 ~ 235 mmHg
Pediatric Mode
SYS 40 ~ 200 mmHg
DIA 10 ~ 150 mmHg
MEAN 20 ~ 165 mmHg
Neonatal Mode
Neonatal Mode
SYS 40 ~ 135 mmHg
DIA 10 ~ 100 mmHg
MEAN 20 ~ 110 mmHg
Resolution
Pressure 1mmHg
Accuracy
Pressure
Maximum Mean error ±5mmHg
Maximum Standard deviation ±8mmHg
Overpressure Protection
Adult Mode 297±3 mmHg
Pediatric Mode 240±3 mmHg
Neonatal Mode 147±3 mmHg
SpO2
Measuring Range 0 ~ 100 %
Alarm Range 0 ~ 100 %
Resolution 1 %
Accuracy 70% ~ 100% ±2 %
0% ~ 69% unspecified
Actualization interval about 1 Sec.
Alarm Delay 10 Sec.
Pulse Rate
Measuring and Alarm Range 0~254bpm
Resolution 1bpm
Accuracy ±2bpm
TEMPERATURE
Channel 1
Measuring and Alarm Range 0 ~ 50 C
Resolution 0.1~C
Accuracy ±0.1~C
Actualization interval about 1 Sec.
Average Time Constant < 10 Sec.
IBP
Label ART PA CVP RAP LAP ICP P1 P2
Measuring and alarm range
ART 0 ~ 300 mmHg
PA -6 ~ 120 mmHg
CVP/RAP/LAP/ICP -10 ~ 40 mmHg
P1/P2 -10 ~ 300 mmHg
Press Sensor
Sensitivity 5 uV/V/mmHg
Impedance 300-3000Ω
Resolution 1 mmHg
Accuracy ±2% or ±1mmHg which great about 1 Sec.
ECG
Lead Mode 5 Leads ( R L F N C or RA LA LL RLV)
Lead selection I II III avR avL avF V
Waveform 2 ch
Lead mode 3 Leads ( R L F or RA LA LL)
Lead selection I II III
Waveform 1 ch
Gain x2.5mm/mV x5.0mm/mV x10mm/mV x20mm/mV auto
HR and Alarm
Range
Adult 15 ~ 300 bpm
Neo/Ped 15 ~ 350 bpm
Accuracy ± 1% or ± 1bpmwhich great
Resolution 1 bpm
Sensitivity > 200 (uV )
Differential Input Impedance > 5 MΩ
CMRR
Monitor > 105 dB
Operation > 105 dB
Diagnosis > 85 dB
Electrode offset potential ±300mV
Leakage Current < 10 uA
baseline Recovery < 3 S After Defi.
ECG Signal Range ±8 m V ( Vp-p )
Bandwidth
Surgery 1 ~ 15 Hz
Monitor 0.5 ~ 35 Hz
Diagnostic 0.05 ~ 100 Hz
Calibration Signal 1 (mV ) Accuracy :±5%
ST Segment Monitoring Range
Measure and Alarm -2.0 ~ +2.0 mV
ARR Detecting
Type ASYSTOLE VFIB/VTAC COUPLET BIGEMINY TRIGEMINY R ON T VT>2 PVC TACHY BRADYMISSED BEATS PNP PNC
Alarm Available
Review Available
RESPARATION
Method Impedance between R-F(RA-LL)
Differential Input Impedance >2.5 MΩ
Measuring Impedance Range: 0.3~5.0Ω
base line Impedance Range: 0 – 2.5 KΩ
Bandwidth 0.3 ~ 2.5 Hz
Resp. Rate
Measuring and Alarm Range
Adult 0 ~ 120 rpm
Neo/Ped 0 ~ 150 rpm
Resolution 1 rpm
Accuracy ±2 rpm
Apean Alarm 10 ~ 40 S
NIBP
Method Oscillometric
Mode Manual Auto STAT
Measuring Interval in AUTO Mode 1 2 3 4 5 10 15 30 60 90 120 180 240480 (Min)
Measuring Period in STAT Mode 5 Min
Pulse Rate Range 40 ~ 240 bpm
Alarm Type SYS DIA MEAN
Measuring and alarm range
Adult Mode
SYS 40 ~ 270 mmHg
DIA 10 ~ 215 mmHg
MEAN 20 ~ 235 mmHg
Pediatric Mode
SYS 40 ~ 200 mmHg
DIA 10 ~ 150 mmHg
MEAN 20 ~ 165 mmHg
Neonatal Mode
Neonatal Mode
SYS 40 ~ 135 mmHg
DIA 10 ~ 100 mmHg
MEAN 20 ~ 110 mmHg
Resolution
Pressure 1mmHg
Accuracy
Pressure
Maximum Mean error ±5mmHg
Maximum Standard deviation ±8mmHg
Overpressure Protection
Adult Mode 297±3 mmHg
Pediatric Mode 240±3 mmHg
Neonatal Mode 147±3 mmHg
SpO2
Measuring Range 0 ~ 100 %
Alarm Range 0 ~ 100 %
Resolution 1 %
Accuracy 70% ~ 100% ±2 %
0% ~ 69% unspecified
Actualization interval about 1 Sec.
Alarm Delay 10 Sec.
Pulse Rate
Measuring and Alarm Range 0~254bpm
Resolution 1bpm
Accuracy ±2bpm
TEMPERATURE
Channel 1
Measuring and Alarm Range 0 ~ 50 C
Resolution 0.1~C
Accuracy ±0.1~C
Actualization interval about 1 Sec.
Average Time Constant < 10 Sec.
IBP
Label ART PA CVP RAP LAP ICP P1 P2
Measuring and alarm range
ART 0 ~ 300 mmHg
PA -6 ~ 120 mmHg
CVP/RAP/LAP/ICP -10 ~ 40 mmHg
P1/P2 -10 ~ 300 mmHg
Press Sensor
Sensitivity 5 uV/V/mmHg
Impedance 300-3000Ω
Resolution 1 mmHg
Accuracy ±2% or ±1mmHg which great about 1 Sec.
Related Accessories
Standard:
ECG Cable (5 Leads) ...... 1Pcs
SpO2 Cable (Adult) ........ 1Pcs
TEMP Probe (Adult) ....... 1Pcs
NIBP Cuff (Adult) ............ 1Pcs
NIBP Tube(150cm) ......... 1PcsGround Cable ................ 1Pcs
Power Cable ................... 1Pcs
Cleaning Cloth ................ 1Pcs
Options:
ECG Cable (3 Leads)
ECG Alligator Clips (metal)
SpO2 Cable (Adult hard shell sensor)
SpO2 Cable (Adult Soft shell sensor)
SpO2 Cable (Neonatal soft sensor)
NIBP Cuff (infant 8 - 14 cm 5/Bag)ECG Alligator Clips (me
SpO2 Cable (Adult hard shell sensor)
SpO2 Cable (Adult Soft shell sensor)
SpO2 Cable (Neonatal soft sensor)
SpO2 Cable (Pediatric hard shell sensor)
SpO2 Cable (Pediatric Soft shell sensor)
SpO2 Cable (Pediatric Soft shell sensor)
SpO2 Cable (Child finger clip)
SpO2 Cable (multi-site Y type)
SpO2 External Cable
NIBP Cuff (Child 13 - 20 cm 5/Bag)
NIBP Cuff (Small Adult 18 - 26 cm 5/Bag)
NIBP Cuff (Adult 26 - 35 cm 5/Bag)
NIBP Cuff (Adult Long 29 - 38 cm 5/Bag)
NIBP Cuff (Large Adult 32 - 42 cm 5/Bag)
NIBP Cuff (Large Adult Long 35 - 44 cm 5/Bag)
NIBP Cuff (Thigh 42 - 50 cm)
NIBP Tube 150cm
NIBP Tube 150cm
Paper Role
Thermal printer
Patient Monitor Wall mounts (3 Models)
Patient Monitor Trolley (5 Models)
IBP Reusable Interface Cable
IBP Disposable Transducers
Thermal printer
Patient Monitor Wall mounts (3 Models)
Patient Monitor Trolley (5 Models)
EtCO2 Module
Water-trap Column
Airway Adapter
IBP Adapter (2-IBP)
CO2 Sample Line
Nasal TubeWater-trap Column
Airway Adapter
IBP Reusable Interface Cable
IBP Disposable Transducers