Product Features

15” high resolution color TFT display with maximum 13 waveforms
Full touch-screen enables intuitive operation by clicking on the specific parameters 
of the waveform in real time standard configuration includes the navigation knob
Pacemaker detection electrosurgical interference proof
Defibrillation protection and defibrillation synchronization
SD Card slot enable memory extension for large data storage
USB and serial port enable abundant future upgrades
VGA output and Analog output
OxyCRG available to judge the respiration and circulation function for neonate
Large font display
Arrhythmia and S-T segment analysis
Built-in rechargeable Li-ion (6hours maximum with 2x 4200mAh batteries)
Nurse Call function and bi-directional communication with MFM-CMS central station
Multi language: English/Simplified Chinese/Traditional Chinese/Spanish/Czech/Italian/Polish/French/German/Brazilian/Portugese/Russian.
Standard configuration: 3/5-lead ECG NIBP SpO2 RESP 2-Temp PR
Optional: 12-lead ECG Nellcor-SpO2 Omron NIBP 2-IBP/4-IBP Mainstream/Side stream CO2 
CO Multi-Gas Full Touch Screen Thermal Recorder WLAN Accessory
Large Data Storage: 120 hours graphic and tabular trends of all parameters; 1200 sets of NIBP
 measurement and 70 alarm events; 2-USB ports and SC card slot ensure data storage/function upgrades.
Clinical data can be viewed and managed via Patientcare viewer software (To be released)
Clinical network: bi-directional communication with CMS-2000 central station by wired or wireless connection
Complete and flexible mounting solution for ICU and operation room. High-end parameters for more flexible clinical applications.


MD9015T Support:
DC power input and to adapt to a variety of environmental conditions
External recorder can print the recorded waveform data and other information in real time
External the EtCO2 interface can be connected to the mainstream CO2 module to facilitate the clinical needs
Drug concentration in titration calculations to support the event functions to customize and record
Up to 32 TCP / IP network connection composed of a central information system
Wireless networking

Options 
Invasive Blood Pressure (IBP)
Carbon dioxide (CO2) Module
Masimo SpO2 Module
Thermal Printer


Product Specifications

Electrocardiogram (ECG)
Input mode Five-lead ECG input / three-lead ECG input
Lead selection I II III aVR aVL aVF V-
Gain selection AUTO0.25X0.5X1X2X4X
Operating mode Diagnostic mode monitoring mode operation mode
Calibration signal 1mV ±5%
Protection The isolation voltage 4000V 50/60Hz. Anti-high-frequency interference of electric knife. Anti-defibrillation protection.
Pacemaker at the same time use Support
Electrode off instructions Each electrode (RL excluded)
Scanning speed 12.5mm/s 25 mm/s 50 mm/s
Common mode rejection ratio ≥90dB
Input impedance ≥5.0MΩ
 
Heartrate (HR)                          

Measuring range Adult 30 ~ 300bpm newborns 30 to 350 bpm
Resolution 1 bpm
Accuracy < ±5bpm
Heart rate sensitivity ≥0.20mVp-p
Alarm range 15 to 350 bpm upper and lower limits
Alarm indication Audible and visual alarm display the data and parameters flashing display alarm 
message text prompts there are three levels of alarm
 
Pulse Oximetry
SPO2 Range:                          1% to 99%
Resolution:                              1%
Accuracy:                                70% to 99% range: ±2%;
1% to 69% range:                    ±5%;
Method:                                  Dual wavelength LED 

ST-segment

ST-segment measurement range -2.0~+2.0mV
ST-segment measurement accuracy Within the range of -0.8 ~ +0.8 mV: ± 0.02mV or reading± 10%
Whichever is greater; Other range: undefined
Resolution 0.01mV
ST-segment update cycle 10s
ST segment alarm range -2.0 ~ +2.0 MV the upper and lower limits continuously adjustable
 
Arrhythmia analysis

Arrhythmia types 13 kinds of (scalable)
 
Respiration Rate
Rate:                                  2 to 120 bpm (CO2)
Resolution:                        1 breath/min
 
NIBP (Non-invasive Blood Pressure)
Technique:                        Oscillometric during inflation
Range:                               Adult/Pediatric: 40 to 240 mmHg
Neonate:                           10 to 150 mmHg
Measurement Cycle:         < 40 sec. typical
Automatic Measurement
Cycles (Selectable):          2 3 5 10 15 30 min; 1 2 4 hr
STAT Mode:                      5 minutes of continuous readings
Max. Allowable Cuff
Pressure: Adult:                300mmHg
Pediatric:                          240 mmHg;
Neonate:                          150 mmHg
Resolution:                       1 mmHg
Transducer Accuracy:       ±2 mmHg or ±2% of reading
Heart (Pulse) Rate
Range:                              30 to 250 bpm
Accuracy:                        ±2 bpm or ±3% whichever is greater


EtCO2 technological index


Interface form External interface
Range of
measurement
0~10 kPa or (0~75) mmHg
precision(error) (0kPa~5.4kPa) the error is ±0.3kPa、
(5.5kPa~10.0kPa)the error is ±0.7kPa
Alarm upper limit 2.1~13.2 kPa
Alarm lower limit 0~6.5 kPa
time 12s
EtCO2 exist
Unit kPa / mmHg / %
Compatibility  
external main stream and side stream are compatible

IBP technological index
 
Measured channel Single way and double way are optional
Range of
measurement
SYS: 6.7kPa~32.0kPa(50mmHg~240mmHg)
MEAN: 3.4kPa~26.6kPa(26mmHg~200mmHg)
DIA: 2.9kPa~24.0kPa(22mmHg~180mmHg)
Mean error ≤±0.4kPa (±3mmHg)
Range of alarm SYS:  upper limit   0.3kPa~53.3kPa(2mmHg~400mmHg)
Lower limit   0kPa~53.1kPa(0mmHg~398mmHg)
MEAN:  upper limit    0.3kPa~53.3kPa (2mmHg~400mmHg)
Lower limit   0kPa~53.1kPa (0mmHg~398mmHg)
DIA:    upper limit    0.3kPa~53.3kPa ( 2mmHg~400mmHg)
          Lower limit   0kPa~53.1kPa ( 0mmHg~398mmHg)
Measured unit mmHg/kPa
Specification of
sensor
DPT-248(space)
Consumables mode SCW-UTAH-01
   
   
   
   
   



Related Accessories

Standard:

ECG Cable (5 Leads) ...... 1Pcs
SpO2 Cable (Adult)  ........ 1Pcs
TEMP Probe (Adult)  ....... 1Pcs
NIBP Cuff (Adult)  ............ 1Pcs
NIBP Tube(150cm)  ......... 1Pcs
Ground Cable   ................ 1Pcs
Power Cable  ................... 1Pcs
Cleaning Cloth  ................ 1Pcs

Options:

ECG Cable (3 Leads)
ECG Alligator Clips (metal)
SpO2 Cable (Adult hard shell sensor)
SpO2 Cable (Adult Soft shell sensor) 
SpO2 Cable (Neonatal soft sensor) 
SpO2 Cable (Pediatric hard shell sensor) 
SpO2 Cable (Pediatric Soft shell sensor)
SpO2 Cable (Child finger clip)
SpO2 Cable (multi-site Y type)
SpO2 External Cable
NIBP Cuff (infant 8 - 14 cm 5/Bag)
NIBP Cuff (Child 13 - 20 cm 5/Bag)
NIBP Cuff (Small Adult 18 - 26 cm 5/Bag)
NIBP Cuff (Adult 26 - 35 cm 5/Bag)
NIBP Cuff (Adult Long 29 - 38 cm 5/Bag)
NIBP Cuff (Large Adult 32 - 42 cm 5/Bag)
NIBP Cuff (Large Adult Long 35 - 44 cm 5/Bag)
NIBP Cuff (Thigh 42 - 50 cm)
NIBP Tube 150cm
Paper Role
Thermal printer
Patient Monitor Wall mounts (3 Models)
Patient Monitor Trolley (5 Models)
EtCO2 Module
CO2 Sample Line
Nasal Tube
Water-trap Column
Airway Adapter
IBP Adapter (2-IBP)
IBP Reusable Interface Cable 
IBP Disposable Transducers